Configuring Optical Elements Information ([Components] Screen); Configuring The Objective Information - Nikon DS-L4 Instructions Manual

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

2

Configuring Optical Elements Information ([COMPONENTS] Screen)

In the [COMPONENTS] screen, you can configure information for optical elements attached to the microscope such as
objectives.

Configuring the Objective Information

2.1
In this section, you configure the information of the objectives attached to the motorized nosepiece or the intelligent
nosepiece. If you have replaced or added objectives, you need to reconfigure the settings.
Prerequisite for configuration
A motorized or intelligent nosepiece must be attached to the microscope in order to perform this configuration.
Registration of calibration settings is required to use DS-L4's scale/measurement function.
To use the DS-L4 scale/measurement function, make sure to register the calibration settings after configuring
objective information. If the [X] button of the [MIC SETUP] screen is pressed after configuring the information of the
objective, you are prompted to register the calibration setting. For the details of the registration of the calibration
settings, see Chapter 11 "11.2.4 Registering the Calibration Settings" in the "Camera Operation" instruction
manual.
Selecting and Configuring the Objective from the List
2.1.1
When appropriate objective is not on the list
When an appropriate objective is not on the list, you can register any objective by pressing the [CUSTOM] button
(only for MA200, L200N/L200ND, L300N/L300ND, LV150A, LV-NCNT2, LV-NCNT-N, and LV100DA-U). For more
information, see "2.1.2 Arbitrarily Configuring the Data of Objective".
[Procedure]
[MIC SETUP] → [COMPONENTS]
(1)
Select the [NOSEPIECE] tab.
The screens in the following procedure are for
LV100DA-U. They vary depending on the model.
(2)
Press the nosepiece address button to be configured.
If the objective is displayed as [-----], information for the
objective is not configured. By default, all addresses are
displayed as [-----].
Chapter 3
Settings
2-14

Advertisement

Table of Contents

Troubleshooting