Configuring the Limit Position of the Motorized XY Stage
6.4.2
[MIC SETUP] → [UTILITY]
(1)
Use the ergo controller or joystick to move the
motorized stage to a point that you want to set as the
limit of movement.
(2)
Press the [SET CURRENT POSITION AS MAX] or [SET
CURRENT POSITION AS MIN] button for X or Y-axis
under [SOFTWARE LIMIT].
The limit position is set based on the current position.
Pressing the [RESTORE DEFAULT] button resets the
setting to the default value.
Maximum/minimum value to be set
When setting the maximum or minimum value, a
value 0.5 m greater or less than the current
position is set (the front, back, right, and left limits
are also the current value added or deducted by
0.5 m).
If the stage is currently at the limit
position, it cannot be returned to default.
In this case, move the stage within the limit
position and then press the [RESTORE DEFAULT]
button once more.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.