Configuring the Interlocked Operation with Switching of Zoom Magnification
5.1.5
The DIA field diaphragm and motorized ND filter wheel can be adjusted automatically when the zoom magnification is
switched.
Prerequisite for configuration
In order to perform this configuration, the motorized DSC zooming port, motorized (or intelligent) nosepiece, and
motorized quadrocular tilting tube must be attached. In addition, the information for objectives must be configured.
To interlock the ND filter, the motorized ND filter wheel must be attached as well. Also, the interlocking function is
disabled when the microscope is in sleep mode.
Device to Be
Setting Item
Interlocked
[F. STOP / ND]
Ni-E Main Body
(DIA field
diaphragm),
motorized ND
filter wheel
[Procedure]
[MIC SETUP] → [MOVEMENT]
Press the [F. STOP/ND] button under [ZOOM INTERLOCK]
to enable or disable the function.
The function is toggled each time you press the button.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.
Chapter 6
Settings
Setting Value
ON
Adjusts the field diaphragm and motorized ND filter
according to the interlock setting for objective
switching. (For the settings, refer to "5.1.1 Configuring
the Interlocked Operation with Switching of
Objectives" in this section. No adjustment is
performed when the interlock setting for objective
switching is set to [OFF].)
OFF (default)
Does not interlock.
1-124
Operation to Be Interlocked