Nikon DS-L4 Instructions Manual page 131

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

Changing the Initial Value of the [INTELLIGENT]
5.1.2
Among the devices interlocked with switching objectives, [INTELLIGENT] can be selected as the operation when
switching the objectives for aperture diaphragm, field diaphragm, or ND wheel. When [INTELLIGENT] is selected, any
adjustment performed during the observation can be reproduced when the objective is returned to the original objective
after switching to another objective. However, once you turn off the DS-L4 and then turn it on, the values of the adjusted
state are cleared and the initial values are restored.
In order to use the [INTELLIGENT] status for aperture diaphragm, field diaphragm, and ND wheel when you turn on the
DS-L4 next or later, you must save the status of the device. This saves the aperture diaphragm, field diaphragm, and ND
wheel status currently used in [INTELLIGENT] to the memory in the microscope and it is used as the initial value of
[INTELLIGENT] when you subsequently turn on the DS-L4.
[Procedure]
[MIC SETUP] → [MOVEMENT]
(1)
Press the [SAVE] button under [OBJECTIVE
INTERLOCK].
A confirmation dialog box appears.
(2)
Press the [OK] button.
The status of the aperture diaphragm, field diaphragm, and
ND wheel, which is used when you select [INTELLIGENT],
is saved.
To cancel saving, press the [CANCEL] button.
When returning the changed initial value of
[INTELLIGENT] to the factory setting
When the information of the objective is reset, the
initial value of [INTELLIGENT] for that nosepiece
address is restored to the factory setting.
See Chapter 6, "2.1 Configuring the Objective
Information" for details on configuring objective
information.
Chapter 6