Configuring the Limit Position for the Elevating Section
6.4.1
You can define a limit position for movement of the elevating section by actually moving it to the desired position.
[Procedure]
[MIC SETUP] → [UTILITY]
(1)
Use the focus knob to move the elevating section to a
point that you want to set as a limit position.
(2)
Press the [SET CURRENT POSITION AS LIMIT] button.
The limit position is defined based on the current position.
Pressing the [RESTORE DEFAULT] button resets the setting
to the default value.
Limit position to be set
For focusing stage system, a limit position is set
0.5 m above the current position and for focusing
nosepiece system, 0.5 m below the current
position.
If the elevating section is currently at the
limit position, it cannot be returned to
default.
In this case, move the elevating section below the
limit position (above for focusing nosepiece
system) and then press the [RESTORE DEFAULT]
button once more.
Z-coordinate for the focusing nosepiece
system
Two systems are used to read Z-axis coordinate
values for the focusing nosepiece system,
application of which depends on the environment
of your microscope system. For details, see "6.4
Setting the Software Limits" in this chapter.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.
Chapter 6
Settings
1-140