■ Operation of the elevating section and motorized XY stage
(Only in Ni-E)
Elevating section can not be
operated using Ni-E main body,
ergo controller, or joystick.
(Only in Ni-E)
Elevating section or motorized
XY stage can not be moved to
an intended position.
The [ESCAPE] button is
disabled.
The [Specimen Removal
Position] button is disabled.
■ Operation of the nosepiece
(Only in Ni-E)
Sometimes, the motorized
nosepiece does not rotate
depending on the position of
the elevating section.
(Only in Ni-E)
Motorized devices are not
interlocked when rotating the
nosepiece, adjusting the zoom
magnification, switching the
optical path, or with the output
of capture command/capture
trigger signal.
(Only in Ni-E)
Interlocked operation when
switching objectives does not
work as intended.
(Only in Ci-E)
Operation of the motorized
swing-out condenser is not
linked to the rotation of the
nosepiece.
The motorized nosepiece skips
addresses.
■ Operation of the dia-illumination lamp/LED
Dia-illumination lamp/LED can
not be turned on or off from
DS-L4.
Chapter 7
Troubleshooting
Operation of the elevating section
is disabled.
Small ranges are set for the
software limit configuration.
The stage is moved to the
specimen removal position.
The elevating section is in the
retracted state.
Rotation stop of the motorized
nosepiece is set to ON.
Interlocking is not configured.
Interlocking is not correctly
configured.
Incorrect configuration of the
objectives.
Interlocking is not configured.
The toggle function is ON.
DS-L4 does not have control.
1-147
Enable the operation.
See Chapter 6, "6.2 Enabling/Disabling the
Operation of the Elevating Section".
Change the software limit setting.
See Chapter 6, "6.4 Setting the Software Limits".
Press the [Specimen Removal Position] button to
reset the specimen removal position.
See Chapter 5, "1.18.3 Retracting the Elevating
Section and Moving the Stage to the Specimen
Removal Position".
Press the [ESCAPE] button to reset the retracted
state of the elevating section.
See Chapter 5, "1.18.2 Retracting the Elevating
Section".
Disable the rotation stop.
See Chapter 6, "5.3 Disabling the Rotation of the
Motorized Nosepiece Depending on the Position
of the Elevating Section".
Configure interlocking.
See Chapter 6, "5.1 Configuring Interlocked
Operation".
Configure correctly.
See Chapter 6, "5.1.1 Configuring the Interlocked
Operation with Switching of Objectives".
Configure correctly.
See Chapter 6, "2.1 Configuring the Objective
Information".
Configure interlocking.
See Chapter 6, "5.6 Configuring Top Lens
Swing-out of the Motorized Swing-out
Condenser".
Turn the toggle function OFF.
See Chapter 6, "5.5 Configuring the Toggle
Function (Switching between Two Objectives)".
Give control to DS-L4.
See Chapter 5, "1.14 Adjusting the
Dia-illumination Lamp/LED".