Nikon DS-L4 Instructions Manual page 137

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

Configuring the Interlocked Operation with Switching of Optical Path
5.1.6
The DIA field diaphragm can be adjusted automatically when the optical path is switched.
Prerequisite for configuration
In order to perform this configuration, the motorized (or intelligent) nosepiece and motorized quadrocular tilting
tube must be attached. In addition, the information for objectives must be configured.
Device to Be
Setting Item
Interlocked
Ni-E Main Body
[F. STOP]
(DIA field
diaphragm)
[Procedure]
[MIC SETUP] → [MOVEMENT]
Press the [F. STOP] button under [PATH INTERLOCK] to
enable or disable the function.
The function is toggled each time you press the button.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.
Chapter 6
Settings
Setting Value
ON
Adjusts the field diaphragm according to the interlock
setting for objective switching. (For the settings, refer
to "5.1.1 Configuring the Interlocked Operation with
Switching of Objectives" in this section. No adjustment
is performed when the interlock setting for objective
switching is set to [OFF].)
OFF (default)
Does not interlock.
1-125