Configuring The Digital Camera Connection; Saving The Settings - Nikon DS-L4 Instructions Manual

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

2

Configuring the Digital Camera Connection

Configure the digital camera connection as follows when connecting the microscope and the camera with a trigger cable
and capturing by outputting a trigger signal from the DSC connector on the microscope.
This configuration is unnecessary if you are capturing with a camera head directly connected to this DS-L4 by pressing the
[CAPTURE] button on the camera control screen.
For details, refer to Chapter 6, "3.1 Configuring the Connection of Digital Camera".
[Procedure]
[MIC SETUP]→[CONNECTION]
(1)
Press the button for the position setting (on the left) for DSC1 or DSC2.
(2)
Select [NOT-CONNECTED] if no camera is connected, otherwise select the position.
(3)
Press the camera manufacturer setting button (on the right) for DSC1 or DSC2.
(4)
Select a camera manufacturer name.
(5)
Repeat steps (1) through (4) when you use two DSC connectors on Ni-E.
Press the [SAVE] button after changing the settings.
After making changes to the settings, be sure to press the [SAVE] button on the [MIC SETUP] - [MAIN] screen to save
the settings to the microscope's memory. If the microscope is turned off without saving settings, previous values are
restored. For details on saving the settings, refer to "3 Saving the Settings" in this chapter.
3

Saving the Settings

After making changes on the [MIC SETUP], be sure to press the [SAVE] button on the [MIC SETUP] - [MENU] screen to
save the settings to the microscope's memory. If the microscope is turned off without saving settings, previous values are
restored.
(1)
After changing a setting, press the [MAIN] button to go back to the [MIC SETUP] - [MAIN] screen.
(2)
Press the [SAVE] button.
A confirmation dialog box appears.
If there is no setting data to be saved, the [SAVE] button is disabled.
(3)
Press the [OK] button.
The settings are saved.
To cancel saving, press the [CANCEL] button.
Pressing the [X] button on the upper-right of the screen closes the [MIC SETUP].
Settings to be saved
Pressing the [SAVE] button will save all changes that have been made since the settings were last saved.
To save the settings individually, perform the above save procedure after each configuration operation.
You do not need to save the settings if you only wish to apply the settings temporarily (i.e. only until the microscope is
turned off).