Configuring the Filter Cube Information (Only for AZ100M, LV100DA-U, SMZ18,
2.2
and SMZ25)
In this section, you configure the information for the filter cube attached to the motorized or intelligent epi-fluorescence
cube turret. If you have replaced or added filter cubes, you need to reconfigure the settings.
Prerequisite for configuration
A motorized or intelligent epi-fluorescence cube turret must be attached to the microscope in order to perform this
configuration.
Selecting and Configuring the Filter Cube from the List
2.2.1
When appropriate filter cube is not on the list
When an appropriate filter cube is not on the list, you can register any name by pressing the [CUSTOM] button
(only in LV100DA-U). For more information, see "2.2.2 Arbitrarily Configuring the Name of Filter Cube (Only for
LV100DA-U)" in this chapter.
[Procedure]
[MIC SETUP] → [COMPONENTS]
(1)
Select the [FL TURRET] tab.
The screens in the following procedure are for
LV100DA-U. They vary depending on the model.
(2)
Press the turret address button to be configured.
In LV100DA-U, address 1 is fixed to BF and address 2 is
fixed to DF. The setting can be modified only for address 3
and 4.
If the filter cube is displayed as [-----], information for the
filter cube is not configured. By default, all modifiable
addresses are displayed as [-----].
(3)
Select the filter cube attached to the microscope and
press the [OK] button.
When [▲] and [▼] buttons are shown at the lower edge of
the list, you can press [▲] or [▼] button to go to other
pages.
The selected filter cube is set to the turret address.
To quit setting, press the [X] button.
(4)
Repeat steps (2) and (3) for all turret addresses to be
configured.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.