Problem
(Ni-E only)
There is no [SLEEP] button on
the [MICROSCOPE
CONTROL] screen.
(Ni-E only)
There is no [SLEEP] button on
the [MIC EASY] screen.
2

Operation

■ General
Problem
Buttons on the microscope
does not work.
(Only in Ni-E)
When the FL SHUTTER button
or CW/CCW button is pressed
on the microscope, the
intended device is not
activated.
(Only in Ni-E)
Operation is not performed as
intended when the buttons on
Ni-E main body or ergo
controller are pressed.
(Only in Ni-E and Ni-U)
Control buttons for nosepiece,
epi-fluorescence cube turret,
etc. are displayed, but do not
work when pressed.
Buzzer does not sound when
buttons are pressed.
The [CW/CCW BUTTON]
selection status is not displayed
on the [BUTTON FUNC] screen
of [MIC SETUP].
Chapter 7
Troubleshooting
Cause
The [SLEEP] button is set to be
hidden on the [BUTTON FUNC]
screen.
The [SLEEP] button is not
configured on the [BUTTON
FUNC] screen to be placed on the
[MIC EASY] screen. Or the
[SLEEP] button is set to be hidden.
Or both.
Cause
Buttons are disabled.
Target for each operation by button
is not correctly configured.
Functions are not correctly
configured.
Intelligent nosepiece or intelligent
epi-fluorescence cube turret is
attached instead of motorized
nosepiece or motorized
epi-fluorescence cube turret.
The device is in sleep state.
Buzzer is turned off.
Settings have been made with Ni
Setup Tool.
1-146
Measure
Configure to show the [SLEEP] button on the
[BUTTON FUNC] screen.
Refer to Chapter 6, "4.1.2 Showing/Hiding the
[SLEEP] Button".
On the [BUTTON FUNC] screen, configure to
place the [SLEEP] button on the [MIC EASY]
screen and shown the [SLEEP] button.
See Chapter 6, "4.1.1 Selecting the Buttons to be
Displayed" and "4.1.2 Showing/Hiding the
[SLEEP] Button".
Measure
Enable the button operation.
See Chapter 6, "4.2.5 Enabling/Disabling the
Button Operation".
Change the targets of button operations.
See Chapter 6, "4.2.2 Changing the Motorized
Shutter Operated with the FL SHUTTER Button"
and "4.2.4 Changing the Motorized Device to be
Operated with CW/CCW Button".
See Chapter 6, "4.2.1 Setting the Function of the
Function Buttons" and "4.3 Configuring the
Functions of the Ergo Controller Buttons".
Intelligent nosepiece or intelligent
epi-fluorescence cube turret can not be operated
from DS-L4. Operate them manually.
Exit the sleep state.
See Chapter 5, "3 Entering the Sleep State (Noise
Reduction)".
Turn on the buzzer.
See Chapter 6, "6.3 Turning ON/OFF the Buzzer".
When you use the Ni Setup Tool, you can select
patterns other than the three types displayed on
the DS-L4 button. In this case, the selection
status is not displayed on DS-L4. (This is not an
error.)