Configuring the Interlocked Operation with Capture Command Sending or
5.1.7
Trigger Signal Output
Motorized quadrocular tilting tube and motorized epi-fluorescence cube turret can be switched by interlocking them with
the sending of a capture command from the microscope's USB connector or output of a capture trigger signal from the
DSC connector.
Prerequisite for configuration
In order to perform this configuration, the target device for interlocking must be attached.
Operation to trigger an interlocked operation
Based on the configuration made here, the specified operation is automatically performed when you send a capture
command/capture trigger signal from the microscope by pressing the [CAPTURE FRONT/LEFT/RIGHT/AUX]
button on the [MICROSCOPE CONTROL] or [MIC EASY] screen or the [CAPTURE] button on the microscope.
Interlocked operation is not performed when you capture the image using the camera head directly connected to
the DS-L4 by pressing the [CAPTURE] button on the camera control screen.
Device to Be
Setting Item
Interlocked
[PATH]
Motorized
quadrocular
tilting tube
[FL TURRET]
Motorized
epi-fluorescence
cube turret
[FL 2nd]
Motorized
epi-fluorescence
cube turret
(2nd layer)
Chapter 6
Settings
Setting Value
Switches the optical path to tube adapter (When
the interlock with optical path switching is enabled,
the field diaphragm is also adjusted according to
[FRONT]
the interlock setting for objective switching. For the
settings, refer to "5.1.1 Configuring the Interlocked
Operation with Switching of Objectives" in this
section.)
Switches the optical path to rear port (When the
interlock with optical path switching is enabled, the
field diaphragm is also adjusted according to the
[REAR]
interlock setting for objective switching. For the
settings, refer to "5.1.1 Configuring the Interlocked
Operation with Switching of Objectives" in this
section.)
[OFF]
Does not interlock.
(Default)
Name of the filter cube
Brings the specified filter cube into the optical path.
at address 1 through 6
[OFF]
Does not interlock.
(Default)
Name of the filter cube
Brings the specified filter cube into the optical path.
at address 1 through 6
[OFF]
Does not interlock.
(Default)