Operation Status Event Register (:Operegister[:Event]) - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

Operation Status Event Register (:OPERegister[:EVENt])

This register hosts the RUN bit (bit 3), the WAIT TRIG bit (bit 5), and the
OVLR bit (bit 11).
• The RUN bit is set whenever the instrument goes from a stop state to a
• The WAIT TRIG bit is set by the Trigger Armed Event Register and
• The OVLR bit is set whenever a 50
• If any of these bits are set, the OPER bit (bit 7) of the Status Byte
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
single or running state.
indicates that the trigger is armed.
Register is set. The Operation Status Event Register is read and cleared
with the :OPERegister[:EVENt]? query. The register output is enabled or
disabled using the mask value supplied with the OPEE command.
Status Reporting
input overload occurs.
9
567

Advertisement

Table of Contents
loading

Table of Contents