5
Commands by Subsystem

:TRIGger:LIN:ID

Command Syntax
:TRIGger:LIN:ID <value>
<value> ::= 7-bit integer in decimal, <nondecimal>, or <string>
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :TRIGger:LIN:ID command defines the LIN identifier searched for in
each CAN message when the LIN trigger mode is set to frame ID.
This command is only valid when the automotive CAN and LIN serial decode option (Option
N O T E
AMS) has been licensed.
Setting the ID to a value of "- 1" results in "0xXX" which is equivalent to
all IDs.
Query Syntax
:TRIGger:LIN:ID?
The :TRIGger:LIN:ID? query returns the current LIN identifier setting.
Return Format
<value><NL>
<value> ::= integer in decimal
Errors
See Also
406
(see
page
586)
from 0-63 or 0x00-0x3f
"- 241, Hardware missing"
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 360
":TRIGger:LIN:TRIGger"
":TRIGger:LIN:SIGNal:DEFinition"
":TRIGger:LIN:SOURce"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
on page 547
on page 354
on page 412
on page 542
on page 409