3
Commands by Subsystem

:TRIGger[:EDGE]:LEVel

Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
<level> ::= 2 V with probe attenuation at 1:1 in NR3 format for
<level> ::= 8 V for digital channels (MSO models)
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15 | EXTernal}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the
active trigger source.
If the optional source is specified and is not the active source, the level on the active source
N O T E
is not affected and the active source is not changed.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current
trigger source.
Return Format
<level><NL>
See Also
366
(see
page
564)
for internal triggers
external triggers
for the MSO models
"Introduction to :TRIGger Commands"
":TRIGger[:EDGE]:SOURce"
":POD<n>:THReshold"
":DIGital<n>:THReshold"
on page 332
on page 369
on page 290
on page 192
Agilent 6000 Series Oscilloscopes Programmer's Reference