7
:ACQuire Commands

:ACQuire:SEGMented:COUNt

Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 25 (w/100K memory) in NR1 format
This command is available when the segmented memory option (Option SGM) is enabled.
N O T E
The :ACQuire:SEGMented:COUNt command sets the number of memory
segments to acquire.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments in the
current acquisition is returned by the :WAVeform:SEGMented:COUNt?
query.
The maximum number of segments may be limited by the memory depth
of your oscilloscope. For example, an oscilloscope with 100K memory
allows a maximum of 25 segments.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 25 (w/100K memory) in NR1 format
See Also
Example Code
190
(see
page
790)
":ACQuire:MODE"
on page 187
":DIGitize"
on page 159
":SINGle"
on page 178
":RUN"
on page 176
":WAVeform:SEGMented:COUNt"
":ACQuire:SEGMented:ANALyze"
"Introduction to :ACQuire Commands"
"Example Code"
on page 191