Replacing the source assembly seals
To avoid excessive leakage of solvent vapor into the laboratory atmosphere,
the following seals must be renewed at intervals of no greater than 1 year:
•
Probe adjuster assembly probe seal
•
Probe adjuster assembly nebulization gas seal
•
Source enclosure seal
•
Nebulizer gas seal
•
Desolvation gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
•
Chemical-resistant, powder-free gloves
•
O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals:
1.
Remove the source enclosure from the instrument (see
2.
Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly (see
•
Probe seal
4-92
Maintenance Procedures
To avoid excessive leakage of toxic and biohazardous
Warning:
solvent vapor into the laboratory atmosphere, the seals listed
below must be renewed, at intervals of no greater than 1 year,
exactly as described in this section.
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.