Removing The Probe Adjuster Assembly Probe Seal And Source Enclosure Seals - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

Probe adjuster assembly nebulization gas seal
Source enclosure seal
Nebulizer gas seal
Desolvation gas seal
4.25.1
Removing the probe adjuster assembly probe seal and source
enclosure seals
4.25.1.1
Required materials
Chemical-resistant, powder-free gloves
O-ring removal kit
Warning:
spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the ion block assembly.
Warning:
open any of the pump inlets while the pump is running.
To remove the probe adjuster assembly probe and source enclosure seals:
1.
Remove the source enclosure from the instrument (see
2.
Use the O-ring removal kit to carefully remove the following seals from the probe adjuster
assembly (see
Probe seal
Nebulizer gas seal
To avoid personal contamination with biohazards or toxic materials, and to avoid
To avoid the pressurized release of oil vapor to the laboratory atmosphere, do not
page
49):
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
September 7, 2016, 715003560IVD Rev. C
page
Page 106
45).