Replacing The Source Assembly Seals; Removing The Probe Adjuster Assembly Probe And Source Enclosure Seals - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Ivd mass spectrometry system
Table of Contents

Advertisement

Replacing the source assembly seals

Warning:
laboratory atmosphere, the seals listed below must be renewed, at
intervals of no greater than 1 year, exactly as described in this
section.
To avoid excessive leakage of solvent vapor into the laboratory atmosphere,
the following seals must be renewed at intervals of no greater than 1 year:
Probe adjuster assembly probe seal
Probe adjuster assembly nebulization gas seal
Source enclosure seal
Nebulizer gas seal
Desolvation gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
Chemical-resistant, powder-free gloves
O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Remove the source enclosure from the instrument (see
2.
Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly:
Probe seal
Nebulizer gas seal
See
"Removing O-rings and seals" on page
To avoid excessive leakage of solvent vapor into the
The source components can be contaminated with
Replacing the source assembly seals
page
4-7).
4-13.
4-87