Focusing the camera:

Adjusting the sprayer tip position

To adjust the sprayer tip position:
1.
Pull open the TRIZAIC source door from the top of the source enclosure.
2.
If you are performing initial installation of a TRIZAIC source, do as
follows:
a.
Remove the spacer from the small space behind the Z control.
b.
Set the X, Y, and Z stages to the zero points on the micrometers.
Reason:
the X, Y, and Z stages set to the extreme positions. The stages must be
reset to the zero point before attempting to tune the position.
The X, Y, and Z controls each have a micrometer scale you can use
Tip:
to position the sprayer tip precisely.
To ensure quality during shipping, the source is shipped with
Camera's
focusing ring
Calibrating the system
2-15