While the Master Alarm is in test mode, the buzzer and the main relay are dis-
abled; however, the auxiliary relay still triggers when an alarm condition occurs.

Configuring Options Available for the Master Alarm

There are three ways you can configure the Master Alarm: front panel keys, RS-
232 serial port, or Web pages.
By using the front panel, you can configure the following options (see table 1-1
on page 1–24).
Table 1-1. Configuring Options Available for the Master Alarm by Using the
Option
User language
Audible return time
N2 address
Clear network
Gas or vacuum type
Alarm message
Alarm system
LED mapping
A Combination Alarm that has a 30-Point Master Alarm with three Area Alarm Modules cannot
interface with a building automation system. The Area Alarm Modules use the same connector that
would otherwise interface with the building automation system. If you need to interface with a
building automation system, order a 30-Point Master Alarm and a Three Module Area Alarm or a
30-Point Master with Six Area Alarm Module Combination Alarm (see table 1-20 on page 1–49).
Page 1-24
Set the language for English, Spanish, or French.
Set the time between audible alarms.
Set the address for the building automation system's notifica-
tion option.
Clears the device list from the Master Alarm's memory.
Set the type of gas or vacuum an input is monitoring.
Set an alarm input message.
Set the system number of the source.
Set the source input to an LED on the front panel.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)