Setup Procedures for Master Alarm

Setting up the Master Alarm involves the following procedures:

Master Alarm Labeling

Tools required:
During the installation of the MedPlus TotalAlert® Alarm Network, a data table
should have been completed for the source input signals that are connected to the
Master Alarm (see table 1-27 on page 1–56).
1. If the input signal data table was not completed during the installation, com-
plete it now by performing the following steps:
a. Record the number of the terminal on the Breakout P.C. Board that each
input is connected to.
b. Record the type of input.
c. Record the LED number that you want to light when that input has an
alarm condition.
NOTE:
Each LED only accepts one type of gas or vacuum.
When you are facing the front panel, the LEDs are numbered one through nine.
1
4
7
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Labeling (see "Master Alarm Labeling" on page 1–55)
Powering on (see "Master Alarm Power-Up" on page 1–60)
Configure by using one the following options:
Front panel keys (see "Configuring the Master Alarm by Using the
Front Panel" on page 1–63)
Web pages (see "Configuring and Browsing the Master Alarm by Using
the Web Pages" on page 1–71)
5/32" hex wrench
Labels (provided)
Input signal data table