Master Alarm Or Area Alarm Transformer; Removal - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Master Alarm or Area Alarm Transformer

Chapter 4: Removal, Replacement, and Adjustment Procedures
4.5

Master Alarm or Area Alarm Transformer

Tools required:

Removal

Turn off the power to the alarm, and leave it off during the installation or
replacement of any component or sub-assembly. Failure to do so could
result in personal injury or equipment damage.
1. Turn off the power to the alarm at the facility's supply.
2. Using a 5/32" hex wrench, unlock the front panel.
NOTE:
For clarity, the front panel is not shown in the illustration for this procedure.
3. Using a phillips head screwdriver, remove the five screws (A) that secure the
cover plates (B) and (C) to the rough-in assembly (D) (see figure 4-6 on page
4–17).
Page 4-16
5/32" hex wrench
Phillips head screwdriver
Needle nose pliers
Shock Hazard:
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)