Local Alarm Buzzer Assembly

Chapter 4: Removal, Replacement, and Adjustment Procedures

4.14 Local Alarm Buzzer Assembly

Tools required:

Removal

1. Using a phillips head screwdriver, loosen the four screws (A) enough to
allow the front panel (B) to slide off (see figure 4-15 on page 4–38).
2. Slide the front panel (B) up and off of the back box (G).
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
3. Turn the power switch (H) to the OFF position.
4. Unplug the buzzer wiring (C).
5. Unplug the power supply wiring (F).
Page 4-38