Serial Number Lost (30-Point Master Alarm with Three Area Alarm Modules)
Chapter 2: Troubleshooting Procedures
2.20 Serial Number Lost (30-Point Master Alarm with Three Area
Alarm Modules)
1. An Area Alarm Module's serial number and the word lost displays on a
30-Point Master Alarm with three Area Alarm Modules.
Yes
2. An Area Alarm Module was replaced or removed from the alarm network.
Yes
3. Do a Clear Network command on the Master Alarm; refer to "Configuring
the Master Alarm by Using the Front Panel" on page 1-63, or "Clear Alarm
Network Device List" on page 1-95. This solves the problem.
Yes
4. Go to "Final Actions" on page 2-7.
Page 2-26
No
Go to "Final Actions" on page 2-7.
No
Ensure all of the Ethernet connections are connected properly.
Refer to the facility's wiring diagram of the Ethernet network. If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, call BeaconMedæs Technical Support.
No
Call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)