No Display on One Area Alarm Module

Chapter 2: Troubleshooting Procedures

2.27 No Display on One Area Alarm Module

1. The displays on the other Area Alarm Modules are ON.
Yes
2. The ribbon cable from the Area Alarm Connector P.C. Board to the Area
Alarm Module is connected.
Yes
3. Replace the ribbon cable (refer to procedure 4.19 on page 4-48). The problem
is solved.
Yes
4. Go to "Final Actions" on page 2-7.

2.28 Audible Signal Does Not Silence

1. Pressing the Silence key cancels the audible signal.
Yes
2. Go to "Final Actions" on page 2-7.
Page 2-36
No
Go to RAP 2.24 on page 2-30. If this solves the problem, go to
"Final Actions" on page 2-7. Otherwise, go to step 2.
No
Connect the ribbon cable from the Area Alarm Connector P.C.
Board to the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to "Final Actions" on page
2-7. Otherwise, go to step 3.
No
Call BeaconMedæs Technical Support.
No
Replace either the Master Alarm Module (refer to procedure 4.2
on page 4-10) or the Area Alarm Module (refer to procedure 4.9
on page 4-26) on the alarm panel that will not silence. If this
solves the problem, go to "Final Actions" on page 2-7.
Otherwise, call BeaconMedæs Technical Support.