Building Automation System (N2) Mapping - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Detailed Component Descriptions
Chapter 1: Introduction
Table 1-3. Browsing Options Available for the Master Alarm by Using the
Browsing Option
Alarms
Event log
Network statistics
Hardware input states

Building Automation System (N2) Mapping

NOTE:
The information in this section is intended to be used by Information Systems
personnel.
The MedPlus TotalAlert® Alarm Network has the option of connecting to a facil-
ity's building automation system (BAS). The BAS accesses the 30 source alarm
points and summary information in the Master Alarm using an N2 open protocol
in the Metasys® Building Management System software. The N2 address is pro-
grammed into the alarm system using the alarm's front panel or the its device
information web page.
The state of each of the 30 source inputs is available using binary input points
BI1 through BI30. Binary input point BI31 is true if there are any active alarms.
For efficient single point monitoring, the best choice is to enroll BI31 in the BAS
polling table.
NOTE:
Change of state (COS) is supported by the software.
The status of each of the 30 source alarm inputs is available using byte points
BYT1 through BYT30. The status of each alarm input returns as follows:
Page 1-26
Lists all of the alarm inputs that communicate
with this Master Alarm.
Log of events and alarms for all inputs that
this Master Alarm is monitoring.
Communication statistics used by Information
Systems personnel.
Lists the state (open or closed) of the inputs.
0 = no alarm
1 = active alarm
2 = silenced alarm
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)