3. Using a phillips head screwdriver, remove the three screws (B) that secure the
cover plate (C) to the rough-in assembly (D).
4. Using needle nose pliers, remove the flag terminals (E) from the power
switch (F).
5. Using needle nose pliers, push in the ears (G) on the power switch (F).
6. Remove the power switch (F).

Replacement

1. To replace the power switch, reverse the order of the removal instructions.
2. To ensure proper operation, perform the "Function Checks" on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Chapter 4: Removal, Replacement, and Adjustment Procedures
Figure 4-18. Power Switch
Power Switch
4
Page 4-45