Ribbon Cables

Chapter 4: Removal, Replacement, and Adjustment Procedures

4.19 Ribbon Cables

Tools required:

Removal

1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-20 on
page 4–48).
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Push open the ears (B) of the connectors holding the ribbon cable (C) that
you want to remove.
4. Remove the ribbon cable (C).
Page 4-48
Phillips head screwdriver
5/32" hex wrench
Figure 4-20. Ribbon Cable
Shock Hazard:
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)