Replacement - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Using a screwdriver, remove the fuse cover (B).
4. Remove the fuse (C).

Replacement

CAUTION:
Replace the fuse with the same type and rating. If the wrong fuse is
installed, equipment may be damaged.
1. Ensure the replacement fuse is the same type and rating as the original fuse.
2. To replace the fuse, reverse the order of the removal instructions.
3. To ensure proper operation, perform the "Function Checks" on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Chapter 4: Removal, Replacement, and Adjustment Procedures
Power Supply Fuse
Page 4-41
4