Transducer - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Transducer

A green LED inside the transducer's transparent housing indicates whether the
transducer is functioning properly (see table 2-4 on page 2-11).
Error Codes
Slow steady flash
Rapid flash
LED off
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Table 2-4. Transducer LED Troubleshooting Codes
Possible Cause
Transducer is functioning
properly.
Device error.
No power to the transducer,
or the transducer has failed.
Error Codes and Troubleshooting Tips
Chapter 2: Troubleshooting Procedures
Corrective Action
No action is required.
Replace the transducer.
Connect the power, or
replace the
2
Page 2-11