Function Checks; Master Alarm Self-Test - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Function Checks

If you are troubleshooting a problem, perform"Initial Actions" on page 2-4
before running the self-tests.

Master Alarm Self-Test

The Master Alarm, Area Alarm, and Local Alarm have self-testing routines.
The Master Alarm routine tests the following:
It also displays the following:
During this routine, the buzzer and the main relay are disabled; however, the aux-
iliary relay still triggers when an alarm condition occurs.
1. To start the test, press and release the Test key.
2. 0 through 9 and A through Z show in rotation for all of the places on the dis-
play.
3. LED Test displays.
4. The LEDs go out.
5. LEDs 1 through 9 individually light green.
6. LEDs 1 through 9 individually light red.
7. Audible Test displays.
8. The alarm sounds for 2 seconds.
9. The flash memory checksum and the Master Alarm's Internet Protocol (IP)
address displays.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Vacuum Fluorescent Display (VFD)
LEDs
Audible alarm
Flash memory checksum
IP address
Programmed alarm setup
LED mapping

Function Checks

Chapter 2: Troubleshooting Procedures
2
Page 2-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents