Replacement - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

6. Remove the front panel (B).
7. Turn the front panel (B) around, and slide it onto two of the four screws (A)
that are used to mount it.
NOTE:
This keeps the front panel out of the way.
8. Unscrew the bottom of the buzzer assembly (D) from the top of the buzzer
assembly (E).
9. Remove the buzzer assembly (D) and (E).

Replacement

1. To replace the buzzer assembly, reverse the order of the removal instructions.
2. To ensure proper operation, perform the "Function Checks" on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Chapter 4: Removal, Replacement, and Adjustment Procedures
Local Alarm Buzzer Assembly
Page 4-39