No Audible Signal On The Local Alarm; Input # Not Assigned - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

No Audible Signal on the Local Alarm

Chapter 2: Troubleshooting Procedures

2.31 No Audible Signal on the Local Alarm

1. The Local Alarm audible indicator does not work.
Yes
2. The wiring from the Local Alarm P.C. board on the back of the front panel to
the buzzer is connected.
Yes
3. The buzzer works properly.
Yes
4. Go to "Final Actions" on page 2-7.

2.32 Input # Not Assigned

1. Input # Not Assigned, displays on the Master Alarm.
Yes
2. The input number shown on the display is configured.
Yes
3. Call BeaconMedæs Technical Support.
Page 2-38
No
Go to "Final Actions" on page 2-7.
No
Connect the buzzer's wiring (refer to procedure 4.14 on page 4-
38). If this solves the problem, go to "Final Actions" on page 2-
7. Otherwise, go to step 3.
No
Replace the buzzer. If this solves the problem, go to "Final
Actions" on page 2-7. Otherwise, call BeaconMedæs Technical
Support.
No
Go to "Final Actions" on page 2-7.
No
Configure the input. See "Configuring the Master Alarm by
Using the Front Panel" on page 1-63. If this solves the problem,
go to "Final Actions" on page 2-7. Otherwise, call
BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)