Area Alarm Labeling

Tools required:
1. If the labeling for the Area Alarm was not completed during installation,
complete it now.
NOTE:
Each Area Alarm Module is factory-labeled with its gas or vacuum type.
2. Using a 5/32" hex wrench, unlock the front panel (CI) (see figure 1-54 on
page 1–109).
Figure 1-54. Area Alarm Labeling (Six Module Area Alarm Shown)
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
5/32" hex wrench
11/32" nut driver
Avery® labels (Avery® part number 8167)
SHOCK HAZARD:

Area Alarm Labeling

Chapter 1: Introduction
Page 1-109