2.4
Transducer Status Fault
1. An Area Alarm Module displays error code A 06.
Yes
2. Transducer is connected properly.
Yes
3. Transducer has a slow steady flash on its heartbeat LED.
Yes
4. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Connect the transducer properly (refer to procedure 4.23 on
page 4-62). If you need more information about how to connect
a transducer, refer to the MedPlus TotalAlert® Alarm Network—
Conversion Kits Installation Instructions. If this solves the prob-
lem, go to "Final Actions" on page 2-7. Otherwise, go to step 3.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 4.
Transducer Status Fault
Chapter 2: Troubleshooting Procedures
2
Page 2-15