Area Alarm Program Errors; Area Alarm Module Ram Error - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Area Alarm Program Errors

Chapter 2: Troubleshooting Procedures
2.8

Area Alarm Program Errors

1. An Area Alarm Module displays error code A 08 or A 09.
Yes
2. Replace the Area Alarm Module (refer to procedure 4.9 on page 4-26). This
solves the problem.
Yes
3. Go to "Final Actions" on page 2-7.
2.9

Area Alarm Module RAM Error

1. An Area Alarm Module displays error code A 04.
Yes
2. Area Alarm Module is connected properly.
Yes
3. Area Alarm Module is working correctly.
Yes
4. If the problem still exists, call BeaconMedæs Technical Support.
Page 2-18
No
Go to "Final Actions" on page 2-7.
No
Call BeaconMedæs Technical Support.
No
Go to "Final Actions" on page 2-7.
No
Connect the Area Alarm Module properly (refer to procedure
4.9 on page 4-26). If this solves the problem, go to "Final
Actions" on page 2-7. Otherwise, go to step 3.
No
Replace the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to "Final Actions" on page
2-7. Otherwise, go to step 4.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)