Area Alarm Breakout P.c. Board; Removal - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Area Alarm Breakout P.C. Board

Chapter 4: Removal, Replacement, and Adjustment Procedures
4.7

Area Alarm Breakout P.C. Board

Tools required:

Removal

1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-8 on page
4–22).
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
2. Turn the power switch (H) to the OFF position.
3. Push open the ears of the connectors (E) to remove the ribbon cable (B).
4. Note the positions of the wiring on the connector (I) for the transducers.
Page 4-22
Phillips head screwdriver
5/32" hex wrench
Needle nose pliers
Figure 4-8. Area Alarm Breakout P.C. Board
Shock Hazard:
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)