Local Alarm P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
5. Unplug the power supply wiring (F).
6. Remove the front panel (B), and save it for installation.
7. Using a 5/16" nut driver, remove the four nuts (D) from the P.C. board (C).
8. Note the positions of the input wiring that run to the P.C. board (C).
9. Remove the input wiring.
10. Remove the P.C. board (C).

Replacement

1. To replace the P.C. board, reverse the order of the removal instructions.
2. To ensure proper operation, perform the "Function Checks" on page 2–5.
4
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-35

Hide quick links: