Start-Up and Checking

Master Alarm Panel

1. Turn on electrical power to alarm
panel.
2. Make following observations:
POWER ON light illuminates on
front of alarm panel.
Boot message is shown on
vacuum fluorescent display.
Booting...
BeaconMedaes
TotalAlert 2 System
Version: [VERSION]
SN 12345
Browse at:
TA2_12345
Alarm modules warm-up for ten
seconds. During warm-up, no
audible or visual alarms activated.
After warm-up, audible and visual
alarms are initiated for any active
alarm.
3. Press MUTE
audible alarm.
4. Make following observations:
If all signals are unprogrammed
(factory default) LEDs will
illuminate green for 10 seconds,
then turn off.
If signals are assigned to LEDs,
multiplexer and LED modules
indicate current status for the
signals and vacuum fluorescent
display displays signal information.
Connected, but unprogrammed
signals are shown on the vacuum
fluorescent display.
Unassigned
Input [INPUT NUMBER]
56
6-847718-00 Rev. B00
http://
button to silence
Status LEDs
Figure 40: Multiplexer/LED Module
NOTE:
If any module appears to be malfunctioning, refer
to Troubleshooting Guide.