Replacement - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Figure 4-6. Master Alarm or Area Alarm Transformer
4. Note the wiring positions of the transformer (E).
5. Using needle nose pliers, remove the flag terminals (F) from the transformer
(E).
6. Using a phillips head screwdriver, remove the two screws (G) that secure the
transformer (E) to the rough-in assembly (D).
7. Remove the transformer (E).

Replacement

WARNING:
Ensure the wiring is correct on the transformer. Failure to do so could
cause personal injury or equipment damage.
1. To replace the transformer, reverse the order of the removal instructions.
2. To ensure proper operation, perform the "Function Checks" on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Master Alarm or Area Alarm Transformer
Chapter 4: Removal, Replacement, and Adjustment Procedures
4
Page 4-17