Area Alarm Module; Removal - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Area Alarm Module

Chapter 4: Removal, Replacement, and Adjustment Procedures
4.9

Area Alarm Module

Tools required:

Removal

If you are replacing an Area Alarm Module in an Area Alarm that uses the
Network Communications Board and the Ethernet, use the Clear
Network function before you start the procedure. Otherwise, the MedPlus
TotalAlert® Alarm Network will indicate that an error has occurred.
If you are replacing an Area Alarm Module in a 30-Point Master Alarm
with three Area Alarm Modules, use the Clear Network function before
you start the procedure. Otherwise, the MedPlus TotalAlert® Alarm
Network will indicate that an error has occurred.
1. If the Area Alarm Module that you are replacing is in an Area Alarm that is
connected to the facility's Ethernet, or in a 30-Point Master Alarm with three
Area Alarm Modules; use the Clear Network function before you begin this
procedure (see "Clear Alarm Network Device List" on page 1–95). Otherwise
the MedPlus TotalAlert® Alarm Network displays the error "Connection Lost
with Device...," because it is searching for the replaced module instead of the
new one.
2. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-10 on
page 4–27).
Page 4-26
Phillips head screwdriver
5/32" hex wrench
CAUTION:
CAUTION:
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)

Hide quick links:

Advertisement

Table of Contents
loading