Browse Gas Pressure Readings, Transducer Locations, Alarm Status, And Error Status - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Browsing the Area Alarm by Using the Web Pages
Chapter 1: Introduction
Browse Gas Pressure Readings, Transducer Locations, Alarm
Status, and Error Status
1. To browse detailed information about the modules, click on the Detailed
Information from Areas link on the Home page. The Area Display Data page
displays (see figure 1-62 on page 1–124).
NOTE:
This information is useful for troubleshooting because it shows the physical loca-
tion of the Network Communications Board and the Area Alarm Modules' read-
ings. It also allows the facility's engineering staff to keep track of all of the pres-
sure readings from the transducers at the same time.
2. Click on the Refresh link to view the most recent information.
Page 1-124
Figure 1-62. Area Display Data—Area Alarm
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)