Master Alarm - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Master Alarm

The Master Alarm displays messages when alarms or errors occur. It always dis-
plays alarms for source equipment. If the Network Communications Board is
installed in the Area Alarm, the Master Alarm displays Area Alarm conditions as
well.
When an alarm occurs the Master Alarm displays the device name when possible.
If the Master Alarm does not know the name of the device, it displays the
device's serial number. Each device has a unique serial number, but each type of
device falls into a serial number range. Knowing the ranges is helpful for trou-
bleshooting alarm messages (see table 2-2 on page 2-9).
Serial Number Ranges
10000 through 19999
20000 through 29999
30000 through 39999
40000 through 49999
50000 through 59999
The factory-programmed alarm messages for the Master Alarm are used to help
identify and solve the problems that may occur with the facility's gas system or
the MedPlus TotalAlert® Alarm Network. There are certain messages available
depending on the type of gas (see table 2-3 on page 2-10).
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Table 2-2. Device Serial Number Ranges
Transducer
Area Alarm Module
Network Communications Board

Master Alarm

Combination Alarm
Error Codes and Troubleshooting Tips
Chapter 2: Troubleshooting Procedures
Device
2
Page 2-9