Transducer Pressure Above Sensor Range

Chapter 2: Troubleshooting Procedures
2.2

Transducer Pressure Above Sensor Range

1. An Area Alarm Module displays error code A 02.
Yes
2. Transducer has a slow steady flash on its heartbeat LED.
Yes
3. If the problem still exists, call BeaconMedæs Technical Support.
2.3

Transducer Communication Error

1. An Area Alarm Module displays error code A 03.
Yes
2. Transducer is connected properly.
Yes
3. Transducer has a steady flash on its heartbeat LED.
Yes
4. If the problem still exists, call BeaconMedæs Technical Support.
Page 2-14
No
Go to "Final Actions" on page 2-7.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
No
Go to "Final Actions" on page 2-7.
No
Connect the transducer properly (refer to procedure 4.23 on
page 4-62). If you need more information about how to connect
the transducer, refer to the MedPlus TotalAlert® Alarm
Network—Conversion Kits Installation Instructions. If this
solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 4.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)