2.23 Data Link Lost with Device: Device Name
1. Data link lost with device: followed by the device name, displays on the
Master Alarm.
Yes
2. The Ethernet connections are connected properly.
Yes
3. The device is powered on.
Yes
4. Call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Ensure all of the Ethernet connections are connected properly.
Refer to the facility's wiring diagram of the Ethernet network. If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
No
Turn the power to the device on. If this solves the problem, go
to "Final Actions" on page 2-7. Otherwise, go to step 4.