2.33 Master Alarm Cannot be Browsed

1. The Master Alarm cannot be browsed.
Yes
2. Press the Test key on the front panel. The address that shows on the display
is a series: 169.254. ##. ##.
Yes
3. The alarm is not finding a server. Type the word Master followed by its serial
number into your Internet browser's address line: for example Master47856.
You still cannot browse the Master Alarm.
Yes
4. Check the LED that indicates whether the alarms are communicating properly
on the Master Alarm Module. See "Master Alarm Power-Up" on page 1-60.
The LED is flashing at a steady rate.
Yes
5. If you still cannot browse the Master Alarm, call BeaconMedæs Technical
Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Ensure that the Ethernet connections are properly connected on
all of the alarm devices (refer to procedure 4.25 on page 4-68).
They are properly connected. If this solves the problem, go to
"Final Actions" on page 2-7. Otherwise, go to step 3.
No
Go to "Final Actions" on page 2-7.
No
Call BeaconMedæs Technical Support.

Master Alarm Cannot be Browsed

Chapter 2: Troubleshooting Procedures
2
Page 2-39