Remote-Mount Transducer; Removal - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Remote-Mount Transducer

Chapter 4: Removal, Replacement, and Adjustment Procedures

4.23 Remote-Mount Transducer

Tools required:

Removal

Do not remove the threaded insert from the transducer. Doing so may
result in attaching the transducer to the incorrect gas or vacuum type,
endangering patient welfare.
1. Leave the threaded insert (H) in the transducer (B). The transducer is not a
serviceable item (see figure 4-24 on page 4–62).
2. Using an adjustable wrench, remove the DISS nut (A) from the DISS demand
valve (I).
Page 4-62
Adjustable wrench
Phillips head screwdriver
Small screwdriver
Leak detection solution
WARNING:
Figure 4-24. Remote-Mount Transducer Assembly
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents