Transducer Wiring Error - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

2.7

Transducer Wiring Error

1. An Area Alarm Module displays error code A 11.
Yes
2. Transducer has a slow steady flash on its heartbeat LED.
Yes
3. Transducer is connected properly.
Yes
4. Check for a pinched transducer wire. Transducer wiring is not pinched.
Yes
5. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
No
Connect the transducer properly (refer to procedure 4.23 on
page 4-62). If this solves the problem, go to "Final Actions" on
page 2-7. Otherwise, go to step 4.
No
Replace the transducer wiring (refer to procedure 4.20 on page
4-50). If this solves the problem, go to "Final Actions" on page
2-7. Otherwise, go to step 5.

Transducer Wiring Error

Chapter 2: Troubleshooting Procedures
2
Page 2-17