OPERATION & MAINTENANCE MANUAL
For Parts or Technical Assistance
Technical Support: 1-888-4MEDGAS (463-3427)
Customer Service: 1-888-4MEDGAS (463-3427)
www.beaconmedaes.com
Area Line
Pressure
Alarm
Product No. 0321
0178-1737-000 Rev. A

Advertisement

Table of Contents
loading

Summary of Contents for BeaconMedaes Area Line Pressure Alarm

  • Page 1 OPERATION & MAINTENANCE MANUAL Area Line Pressure Alarm Product No. 0321 For Parts or Technical Assistance 0178-1737-000 Rev. A Technical Support: 1-888-4MEDGAS (463-3427) Customer Service: 1-888-4MEDGAS (463-3427) www.beaconmedaes.com...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS HEADER User Responsibility ......1 User Responsibility Cross Connection Prevention ....1 This Product will perform in conformity with the descrip- Precautions .
  • Page 3: 1/Introduction

    2/GENERAL INFORMATION Dimensions 2.1 Description Alarm Panel: The area line pressure alarm panel is a wall mounted unit 14 in. x 9 3/4 in. (35.6cm x 24.6cm) that monitors the gas pressure (or vacuum) in up to six separate lines and sounds an alarm if preset limits are Enclosure (rough-in box): violated (Figure 1).
  • Page 4 2/GENERAL INFORMATION HEADER Figure 2 Six Gas Area Alarm Panel Switch Gauge Operation (NORMAL): The gauge pointer moves in response to change in gas line pressure. Under normal pressure conditions, the pointer will not be in con- tact with the gauge alarm contacts. Switch Gauge Operation (ABNORMAL): As gas line pressure goes to an abnormal (high or low) level, the pointer will move accordingly.
  • Page 5: Pre-Use Checkout

    2/SETUP AND OPERATION HEADER 3.1 Pre-Use Checkout 3.2 Controls and Indicators WARNING: After installation of an alarm panel, make (Figure 5) absolutely certain that the alarms are correctly con- Alarm Silence - Press this button to silence the alarm. nected to their respective gas lines. The alarm remains silenced unless a new alarm condition After the builder has installed the alarm panels, they must occurs.
  • Page 6: 4/Operator Maintenance

    4/OPERATOR MAINTENANCE HEADER Contact BeaconMedæs Technical Support for service 4.1 Repair Policy assistance. In all cases, other than where BeaconMedæs’s warranty is applicable, repairs will be Note: Service, other than what is described in this manu- made at BeaconMedæs’s current list price for the replace- al, must be performed by a “Technically Competent”...
  • Page 7: Testing Switch Gauges

    4/OPERATOR MAINTENANCE HEADER 4. Mark the exact location of the contact assembly’s 4.3 Testing the Switch Gauges brass pins on the transparent tape (Figure 8B). Use the following procedure to test the switch gauge cir- 5. Using the 1/16 hex wrench individually turn each low cuitry.
  • Page 8 4/OPERATOR MAINTENANCE HEADER Note: Each line pressure or vacuum gauge has a corre- Figure 7 sponding switch gauge. Gauge Location Note: For clarity, shorting bars are not shown. Figure 8 Pressure Switch Testing 0178-1737-000 Rev. A...
  • Page 9: Warranty

    WARRANTY BeaconMedæs warrants the Area Line or inadequately maintained, field modi- Pressure Alarm to be free of defects in fied in any way, improperly repaired, or materials or workmanship when installed in any other way improperly applied or and operated in accordance with instruc- used.
  • Page 10 Part No. 0178-1737-000 Rev. A ©2004 BeaconMedæs. All rights reserved.