Area Alarm Power-Up - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Area Alarm Power-Up

Chapter 1: Introduction

Area Alarm Power-Up

Tools required:
1. Using a 5/32" hex wrench, unlock the front panel (DF) (see figure 1-55 on
page 1–112).
2. Turn the power switch (DG) to the ON position.
For three seconds, the Area Alarm display (BH) shows
init.
Then it displays the appropriate gas or vacuum pressure from the transducers.
3. Locate the P.C. board (DH).
4. Determine whether the P.C. board is the Network Communications Board
(DH) or Connector P.C. Board (not shown).
NOTE:
The Connector P.C. Board does not have LEDs on it.
Page 1-112
5/32" hex wrench
Figure 1-55. Area Alarm
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)